Available signals
• Stimulus: pure tone, warble tone
• 2 external inputs and USB audio for speech audiometry
• MIC input for live speech audiometry
• Masking: NBN, WN, SN
Signals specification
• Attenuator step: 1, 2 and 5 dB
• Presentation: Continuous, Pulsed (0.5, 1 and 2 Hz)
• Warble: 5 Hz sin wave modulating signal
Available outputs and transducers
• AC: Headphones (TDH-39 / DD45)
• HF headphones (HDA-300)
• Insert earphones (ER-3 / ER-5)
• BC: B-71 bone vibrator
• Free field
• Insert masking earphone: IME-100
Available tests
• Pure Tone audiometry (HL and UCL)
• Auto threshold (modified Hughsone-Westlake)
• HF audiometry: from 8 to 20 kHz (optional)
• Speech audiometry (SRT, WRS, MCL, and UCL)
• SISI test
• Stenger, with pure tone or speech stimulation
• 2 independent channels Master Hearing Aid
• QuickSIN® test (optional)
• Visual Reinforcement Audiometry (VRA) and Conditioned Play Audiometry (CPA) test (optional) with video reinforces
Patient - operator communication
• Talk over: external microphone included
• Talk back: through monitor headset (included); patient microphone included
• Patient response trigger
Monitor signal
• Both channels monitored through the monitor headset (included)
Calibration
• Validity: 12 months
• Separated calibration tables for each type of transducer (AC, AC-INS, AC-HF, INS-MASK, BC, FF)
• All the parameters stored in the device through a dedicated software
Computer interface
• Connection: USB (driverless)
• Compatible software: - Inventis Maestro
Minimum computre requirements
• Processor: Intel® Core 2 Duo, 1.6 GHz
• RAM: 4 GB
• Operative system: Windows 7-8-10, 32 or 64 bits
Power supply
• Maximum consumption: 10 Watts
• Power supply: 6V, 2A cont., through an external medical grade 100-240 Vac 50/60 Hz power supply
Mechanics
• Size (WxDxH): 30.8 × 21.7 × 6.7 cm / 12.1 × 8.5 × 2.5 in
• Weight: 1.3 kg / 2.9 lbs