Tecno-gaz - Classe B - 15 litres
Tecno-gaz - Classe B - 15 litres
• Operating voltage: 230V-50/60 Hz
• External size mm: 510×590×390
• Weight Kg: 54
• Max absorbed power: 1.700 W
• Chamber size mm: Ø 245×320
Sterilization cycles
121° C wrapped and unwrapped
• Exposed time: 18 min.
• Drying time: 17 min.
• Press. work min and max (relative bar): 1.04 ÷ 1.30
• Temperature work min and max: 121 ÷ 125
134° C wrapped and unwrapped
• Exposed time: 5 min.
• Drying time: 17 min.
• Press. work min and max (relative bar): 2.02 ÷ 2.38
• Temperature work min and max: 134 ÷ 138
134° C FLASH
• Exposed time: 4 min.
• Drying time: 2 min.
• Press. work min and max (relative bar): 2.02 ÷ 2.38
• Temperature work min and max: 134 ÷ 138
134° C PRION
• Exposed time: 20 min.
• Drying time: 17 min.
• Press. work min and max (relative bar): 2.02 ÷ 2.38
• Temperature work min and max: 134 ÷ 138
134° C HELIX/ BOWIE&DICK
• Exposed time: 3.5 min.
• Drying time: 4 min.
• Press. work min and max (relative bar): 2.02 ÷ 2.38
• Temperature work min and max: 134 ÷ 138
VACUUM
• Press. work min and max (relative bar): - 0.9
The pre-heating time and fractionation of the vacuum can vary depending on the conditions of the autoclave from 25 - 35 minutes that will add up with the times of the cycles shown in table. All sterilization cycles have 3 stages of vacuum.
Sterilization cycles
121° C wrapped and unwrapped
• Exposed time: 18 min.
• Drying time: 17 min.
• Press. work min and max (relative bar): 1.04 ÷ 1.30
• Temperature work min and max: 121 ÷ 125
134° C wrapped and unwrapped
• Exposed time: 5 min.
• Drying time: 17 min.
• Press. work min and max (relative bar): 2.02 ÷ 2.38
• Temperature work min and max: 134 ÷ 138
134° C FLASH
• Exposed time: 4 min.
• Drying time: 2 min.
• Press. work min and max (relative bar): 2.02 ÷ 2.38
• Temperature work min and max: 134 ÷ 138
134° C PRION
• Exposed time: 20 min.
• Drying time: 17 min.
• Press. work min and max (relative bar): 2.02 ÷ 2.38
• Temperature work min and max: 134 ÷ 138
134° C HELIX/ BOWIE&DICK
• Exposed time: 3.5 min.
• Drying time: 4 min.
• Press. work min and max (relative bar): 2.02 ÷ 2.38
• Temperature work min and max: 134 ÷ 138
VACUUM
• Press. work min and max (relative bar): - 0.9
The pre-heating time and fractionation of the vacuum can vary depending on the conditions of the autoclave from 25 - 35 minutes that will add up with the times of the cycles shown in table. All sterilization cycles have 3 stages of vacuum.